ORIGINAL RESEARCH ARTICLE

Capability of inversely planned intensity-modulated radiotherapy delivery in Telecobalt Bhabhatron 3i Machine

Ramamoorthy Ravichandran1* Tarani Mondal1 Manikandan Palanisamy2 Nandhiya Muthuraman2 G.V. Subrahmanyam2
Show Less
1 Department of Radiation Oncology, Medical Physics Unit, Cachar Cancer Hospital and Research Centre, Silchar, Assam , India
2 Panacea Medical Technologies Pvt Ltd, Malur, Karnataka , India
Received: 2 June 2026 | Revised: 12 August 2026 | Accepted: 26 August 2026 | Published online: 14 September 2026
© 2026 by the Author(s). This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution 4.0 International License ( https://creativecommons.org/licenses/by/4.0/ )
Abstract

The Bhabhatron 3i (Bh 3i) is a higher-specification O-ring telecobalt machine manufactured in India. This machine has built-in multileaf collimator (MLC) vanes with high-speed motors for executing dynamic inverse planned intensity-modulated radiotherapy (IP-IMRT). This machine’s ability to deliver IP-IMRT is validated. Dosimetric measurements carried out in Bh 3i were compared to the TrueBeam Varian linac, available in another hospital. Measured radiation field analyzer data of Bh 3i were configured in the treatment planning system (TPS) on a request basis, and a temporary license was obtained from a TPS manufacturer, M/s Prowess, USA. Quality assurance measurements were carried out on the planned dose distribution for test IP-IMRT plans, downloaded from the American Association of Physicists in Medicine Task Group 119 (TG 119). MLC leakage value in Bh 3i was much lower than that of the TrueBeam linac. MLC settings remained constant in all gantry orientations. The dynamic dose delivery patterns in Bh 3i and the TrueBeam linear accelerator were similar. Static forward-planned multiple MLC field IMRT with the same isocentre produced good treatment plans. Dynamic IMRT planned on TG 119 test cases of tumor volumes demonstrated satisfactory conformality and homogeneity. Doses to organs at risk were acceptable as per norms. It was found that treatment plan goals are achievable in Bh 3i IP- IMRT, similar to the TrueBeam linac. The new Bh 3i cobalt machine can deliver IP-IMRT for cancer treatment. Obtaining a permanent license for the required TPS software is a high priority to secure the further clearances needed for clinical use.

Keywords
Bhabhatron 3i
Intensity-modulated radiotherapy
Multileaf collimator leakage
Conformal radiotherapy
Inverse planning
Funding
None.
Conflict of interest
The authors declare that there are no competing interests with any agencies, as well as among themselves.
References
  1. Page BR, Hudson AD, Brown DW, et al. Cobalt, linac, or other: What is the best solution for radiotherapy in developing countries? Int J Radiat Oncol Biol Phys. 2014;89(3):476-480. doi: 10.1016/j.ijrobp.2013.12.022
  2. Van Dyk J, Battista JJ, Almond PR. A retrospective of Co-60 radiotherapy: The atom bomb that saves lives. Med Phys Int J. 2020;4(Spl Issue. History of Medical Physics):327-350. Accessed September 11, 2026. https://amos3.aapm.org/abstracts/pdf/166-58831-15631646-171798-1721147678.pdf
  3. Nagaraj H, Bhudatt Paliwal, et al. Debate on clinical efficacy and future projection on applications of telecobalt machine. J Cancer Res Ther. 2014;10:781-790.
  4. Chu JCH, Coia LR, Aziz D, Stafford PM. Dose to superficial node for patients with head and neck cancer treated with 6 MV and 60Co photons. Radiother Oncol. 1991;21(4):257-260. doi: 10.1016/0167-8140(91)90050-Q
  5. Kannan R, Dutta G, Ravichandran R, et al. Feasibility and acute toxicity of seven fractions per week accelerated radiotherapy with or without concurrent chemotherapy in laryngopharyngeal squamous cell carcinoma: A pilot study. J Head Neck Physicians Surg. 2026;14(1):86-92. doi: 10.4103/jhnps.jhnps_5_26
  6. Van Dyk J, Battista JJ. Co-60 an old modality, a renewed challenge. Curr Oncol. 1996;3(10):8-17. Accessed September 11, 2026. https://www.researchgate.net/publication/238082452_Cobalt60_An_Old_Modality_A_Renewed_Challenge
  7. Joshi C, Dhanesar S, Darko J, Kerr A, Vidyasagar P, Schreiner LJ. Practical and clinical considerations in Co-60 tomotherapy. J Med Phys. 2009;34(3):137. doi: 10.4103/0971-6203.54847
  8. Jayarajan K, Kar DC, Sahu R, Radke MG, Singh M. BARC develops Co-60 teletherapy machine for cancer treatment. BARC Newsletter. 2005;253:10-14. Accessed September 11, 2026. https://barc.gov.in/barc_nl/2005/200502.pdf
  9. Singh IR, Ravindran PB, Ayyangar KM. Design and development of motorized multileaf collimator for telecobalt unit. Technol Cancer Res Treat. 2006;5(6):597-605. doi: 10.1177/153303460600500607
  10. Adams EJ, Warrington AP. A comparison between cobalt and linear accelerator-based treatment plans for conformal and intensity modulated radiotherapy. Br J Radiol. 2008;81(964):304-310. doi: 10.1259/bjr/77023750
  11. Schreiner LJ, Joshi CP, Darko J, Kerr A, Salomons G, Dhanesar S. The role of Co-60 in modern radiotherapy dose delivery and image guidance. J Med Phys. 2009;34(3):133-136. doi: 10.4103/0971-6203.54846
  12. Macomber MW, Fong N, Sponsellar PA, et al. A novel compensator device to deliver IMRT with Co-60 units in low- and middle-income countries: A comparative dosimetric analysis. Int J Radiat Oncol Biol Phys. 2017;99(2):S231. doi: 10.1016/j.ijrobp.2017.06.567
  13. Oh K, Sengupta B, Olanrewaju A, et al. Commissioning and dosimetric validation of a novel compensator based Co-60 IMRT system for evaluating suitability to automated treatment planning. Med Phys. 2023;50(7):4466-4479. doi: 10.1002/mp.16423
  14. Li HH, Rodriguez VL, Green OL, et al. Patient specific quality assurance for the delivery of (60)Co intensity modulated radiotherapy subject to a 0.35 T lateral magnetic field. Int J Radiat Oncol Biol Phys. 2015;91:65-72. doi: 10.1016/j.ijrobp.2014.09.008
  15. Wooten HO, Rodriguez VL, Green OL, et al. Bench mark IMRT evaluation of a Co-60 MRI guided radiotherapy system. Radiother Oncol. 2015;114(3):402-405. doi: 10.1016/j.radonc.2015.01.015
  16. Sahani G, Kumar M, Sharma PD, et al. Compliance of Bhabhatron-II telecobalt unit with IEC standard - Radiation safety. J Appl Clin Med Phys. 2009;10(2):120-130. doi: 10.1120/jacmp.v10i2.2963
  17. Sharma SD, Kumar R, Kakade NR, et al. Development of a MulriLeaf Collimator for indigenous Telecobalt machine Bhabhatron-II. In: Proceedings of the 37th Annual Conference of Association of Medical Physicists of India. Hyderabad, India; 2016:I-07:8-9.
  18. Govinda Rajan KN. Project introduction and a compensator based IMRT for cobalt. J Med Phys. 2017;42(suppl 1):21. Accessed September 11, 2026. https://ampi.org.in/wp-content/uploads/2018/10/2017-JMP-Suplement-AMPICON2017.pdf
  19. IMRT Treatment Planning-Forward Planning. In: Van Dyk J, ed. The Modern Technology of Radiation Oncology. Vol 2. Madison, Wisconsin, USA: Medical Physics Publishing; 1999. Chapter 6, page 223.
  20. Murthy KK, Sivakumar SS, Davis CA, Ravichandran R, El Ghamrawy K. Optimization of dose distribution with multi-leaf collimator using field-in-field technique for parallel opposing tangential beams of breast cancers. J Med Phys. 2008;33(2):60-63. doi: 10.4103/0971-6203.41194
  21. Al-Rahbi ZS, Al Mandhari Z, Ravichandran R, et al. Dosimetric comparison of intensity modulated radiotherapy isocentric field plans and field in field (FIF) forward plans in the treatment of breast cancer. J Med Phys. 2013;38(1):22-29. doi: 10.4103/0971-6203.106601
  22. Rani AR, Ayyangar K, Reddy AR, Kumar AA, Reddy PRY. Efficacy of MLC based Co-60 plans: A DVH comparison and analysis with 6 MV plans. Med Dosim. 2021;46(1):80-85. doi: 10.1016/j.meddos.2020.09.002
  23. Intensity Modulation Methods. In: Van Dyk J, ed. Beam Shaping and Intensity Modulation. The Modern Technology of Radiation Oncology. Madison, Wisconsin, USA: Medical Physics Publishing; 1999. Chapter 12:466-472.
  24. Ravichandran R, Bonanthaya RK, Mani T, et al. A new model O Ring isocentric telecobalt machine with built in imaging system and multi-leaf beam collimation. Int J Radiol Radiat Ther. 2024;11(5):129-135. doi: 10.15406/ijrrt.2024.11.00400
  25. Ravichandran R, Subrahmanyam GV. An Elite Version of Telecobalt Machine with O-ring Design for Clinical Radiotherapy. J Med Phys. 2024;49(4):719-720. doi: 10.4103/jmp.jmp_164_24
  26. Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359-5373. doi: 10.1118/1.3238104
  27. Ravichandran R. Optimization of radiotherapy resources in cancer treatments: Issues relating to cobalt treatments in Indian Context. Int J Radiol Radiat Ther. 2026;13(2):58-62. doi: 10.15406/ijrrt.2026.13.00455
Share
Back to top
Advances in Radiotherapy & Nuclear Medicine, Electronic ISSN: 2972-4392 Print ISSN: 3060-8554, Published by AccScience Publishing